SPECIFICATIONS
17
Electrical Characteristics
Service required: 2 Pole, 3 Wire (single phase plus ground)
Compressor P/N
Customer's Electrical
Service
Compressor
RLA
Requires Setting of
Transformer Voltage Taps
267794D26G or
268086D26G
220 V~ (
±
10%), 60 Hz
200 V~ (
±
5%), 50 Hz
16 A
NO
268613D26G
220 V~ (
±
10%), 60 Hz
200 V~ (
±
5%), 50 Hz
8 A
NO
267794D1526G or
268086D1526G
220 V~ (
±
10%), 60 Hz
200 V~ (
±
5%), 50 Hz
220 V~ (
±
5%), 50 Hz
16 A
YES
268613D1526G
220 V~ (
±
10%), 60 Hz
200 V~ (
±
5%), 50 Hz
220 V~ (
±
5%), 50 Hz
8 A
YES
267794D1626G or
268086D1626G
220 V~ (
±
10%), 60 Hz
200 V~ (
±
5%), 50 Hz
230/240 V~ (
±
5%), 50 Hz
16 A
YES
268613D1626G
220 V~ (
±
10%), 60 Hz
200 V~ (
±
5%), 50 Hz
230/240 V~ (
±
5%), 50 Hz
8 A
YES
267794D2526G or
268086D2526G
220 V~ (
±
10%), 60 Hz
200 V~ (
±
5%), 60 Hz
17.5 A
YES
268613D2526G
220 V~ (
±
10%), 60 Hz
200 V~ (
±
5%), 60 Hz
8 A
YES
NOTE
Transformer voltage taps must be selected to match the incoming supply voltage.
See section on Mains Power Supply Connection.
Power required: 3.4 kW, 16.0 amperes full load at 220 V~, 60 Hz and
3.0 kW, 15.5 amperes full load at 200 V~, 50 Hz.
60 amperes starting current, 73 amperes locked-rotor current.
Power Cord Connector: NEMA L6-20P (2 pole, 3 wire, 20 A, 250 V~) male plug
to connect to customer’s receptacle.
Power Conductor Requirements for hard wiring (if not using the compressor power cord):
Use 12 AWG (3.3 mm
2
) copper wire rated at 250 V~ minimum, protected with oil-
resistant insulation. Two power leads and one grounding conductor are required.
Power should be provided through a separately fused safety switch with 20-ampere fuses.
Fuses are SLO-BLO, type T. Fuse ratings are 250 volts, 1.0 ampere, 0.63 ampere, 0.4 ampere
and 5.0 amperes. See the Description and the Parts sections.
Cold Head Receptacle: Receptacle
on the
compressor front panel used with a cold head cable
provides power to the cold head.
Accessory Receptacle on the compressor used with an optional accessory cable provides
remote on/off switching capability.
Summary of Contents for HC-4A
Page 2: ......
Page 6: ...iv This page is intentionally blank ...
Page 10: ...4 This page is intentionally blank ...
Page 12: ...6 This page is intentionally blank ...
Page 14: ...8 This page is intentionally blank ...
Page 36: ...30 This page is intentionally blank ...
Page 38: ...32 This page is intentionally blank ...
Page 59: ...Troubleshooting 53 Figure 8 HC 4A Wiring Diagram Main Electrical Chassis and Components ...
Page 60: ...Troubleshooting 54 Figure 9 HC 4A2 Wiring Diagram Main Electrical Chassis and Components ...
Page 61: ...Troubleshooting 55 Figure 10 HC 4A and HC 4A2 Wiring Diagram Air Cooled Unit ...
Page 62: ...Troubleshooting 56 Figure 11 HC 4A Electrical Schematic ...
Page 64: ...Troubleshooting 58 Figure 13 HC 4A2 Electrical Schematic Air Cooled Unit ...
Page 68: ...Parts 62 Figure 15 Parts Identification for HC 4A 12 14 19 23 ...
Page 69: ...Parts 63 Figure 16 Parts Identification for HC 4A2 12 14 19 23 ...
Page 70: ...Parts 64 Figure 17 Parts Identification 29 28 30 31 ...