INSTALLATION
21
The HC-4A Zephyr
®
Air-Cooled Compressor is shipped in an operable condition. Installation
consists of:
Unpacking
Examining for damage
Removing shipping bolt
Placing the compressor at its point of use
Connecting utilities
Checking compressor operation
Connecting other system components.
Introduction
Install the Compressor and the Gas Lines according to the following procedures.
To prevent contaminating the components or the system, it is important to follow the procedures
in this manual step by step.
NOTE
Be sure to have 99.995% pure helium gas available at installation. See
Refrigerant Quality in Specifications.
Receipt Inspection Instructions
1.
Upon
receipt, inspect the shipping container and the compressor for
damage.
AVOID EQUIPMENT DAMAGE.
Remove the shipping bolt retaining the
compressor motor before running the compressor.
AVOID CONTAMINATION.
When checking the compressor for shipping
damage do not connect gas lines and cold head. The components may become
contaminated with compressor oil.
AVOID EQUIPMENT FAILURE, CONTAMINATION OR NUISANCE
SHUTDOWN.
Do not tip the compressor more than 5 degrees from horizontal to
avoid flowing oil into unwanted places.
1.1.
If there is any evidence of external damage to the container, be sure the carrier’s driver
sees the damage. Note it on the shipping documents and have the driver acknowledge
it by his initials on the delivery receipt.
1.2.
Remove the compressor from its shipping container and inspect for damage. If there
was external damage to the compressor, remove its covers and check for internal
damage. Notify the carrier immediately and take photographs of the damage to
document your claim to the carrier. Keep the damaged shipping container.
Summary of Contents for HC-4A
Page 2: ......
Page 6: ...iv This page is intentionally blank ...
Page 10: ...4 This page is intentionally blank ...
Page 12: ...6 This page is intentionally blank ...
Page 14: ...8 This page is intentionally blank ...
Page 36: ...30 This page is intentionally blank ...
Page 38: ...32 This page is intentionally blank ...
Page 59: ...Troubleshooting 53 Figure 8 HC 4A Wiring Diagram Main Electrical Chassis and Components ...
Page 60: ...Troubleshooting 54 Figure 9 HC 4A2 Wiring Diagram Main Electrical Chassis and Components ...
Page 61: ...Troubleshooting 55 Figure 10 HC 4A and HC 4A2 Wiring Diagram Air Cooled Unit ...
Page 62: ...Troubleshooting 56 Figure 11 HC 4A Electrical Schematic ...
Page 64: ...Troubleshooting 58 Figure 13 HC 4A2 Electrical Schematic Air Cooled Unit ...
Page 68: ...Parts 62 Figure 15 Parts Identification for HC 4A 12 14 19 23 ...
Page 69: ...Parts 63 Figure 16 Parts Identification for HC 4A2 12 14 19 23 ...
Page 70: ...Parts 64 Figure 17 Parts Identification 29 28 30 31 ...