OPERATING MANUAL FOR SELECT EDI 60
L300180 Iss:A
39
11.3
Cartridge media pack
Your
Select edi 60 is supplied with a Pretreat PT8 cartridge pack and a
blank ‘Flushing pack’
Note: The PT8 pack must be fitted in the left hand position in the front of
the unit and the Flushing pack should be fitted in the right hand position.
Refer to Section 9, Consumables, for part nos. of all consummable items.
11.4
Unit positioning
The
Select edi 60 can be either bench or wall mounted.
11.4.1
Bench mounted
Locate the unit on a suitable work surface, ensuring access to tank
dispense bib tap and sufficient clearance to rear of the machine for all
water and electrical connections.
All water feed connections are found on the rear of the unit. Using the
installation kit provided attach the water feed, drain and overflow tubing
as shown below.