OPERATING MANUAL FOR SELECT EDI 60
L300180 Iss:A
20
“Line Cell Constants”
There are three line cells within the unit, each cell constant can be set
independently and are displayed. The accepted input range will be 0.05
– 1.99. If a value outside of this range is entered, the value will be
ignored, and the previous value retained.
“Transit Mode”
When the unit is first delivered it will be set in “Transit mode”, When the
unit is first powered up the user will be prompted to set the time & date
and reset the consumable dates. After the time & date options have
been entered the Transit mode option will automatically go to
“Disabled”
“Touchscreen Calibration”
The calibration of the touch screen will be able to be carried out within
this menu option. A number of identified calibration points will be
highlighted
on
the
touch
screen.
Pressing
the
highlighted
points/markers will cause the resulting values to be stored and used in
place of the default values. Default values will be entered/used for
Touch screen calibration whenever
Commission Date
This feature when selected will display the date on which the unit was
commissioned. The date is automatically entered following the setting
of the time and date after Transit Mode. If the unit is re-commissioned
at any point in the future the date can be changed within this menu.
The default date is set as 01/01/09
Software Update
This menu option will provide a means of uploading/replacing software
on the unit’s control and display boards. Files will be able to be
uploaded direct from a USB memory device, which can be plugged into
the display board/pod. Three sub menus will be displayed on selecting
this feature.
“Update Display”
“Update Controller”
“Update GUI”
When any one of these options is selected, an “Are You Sure”
message will be displayed. If “Yes” is selected the unit will upload the
appropriate files from the USB memory device. The unit’s software will