CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018
68 of 112
DLM 97200-03 EN Rev. A
2. Use the horizontal scroll bar to move the cursor between the hour, minute, and second fields,
and use the vertical scroll bars to set the value.
N
OTE
: When using the CheckPoint Portable/On-Line TOC Sensor in applications with high-
temperature samples (>60 °C), the Sensor should be operated in On-Line Average mode with a
measurement output rate of at least 5 minutes.
To change the measurement output rate, press the
button, select
Setup
Output rate
Averaged
, and then use the arrow keys to set the output rate.
Password
You can set a password that limits access to various screens.
1. On the
Setup
screen, use the
Vertical Scroll
bar to highlight
Password
, and then press the
Enter
key.
2. Press
F1
to toggle the password state between
Enabled
and
Disabled
. When you enable
the password, you will be prompted to enter a password, as described in
3. Press
F2
to change the password.
4. To specify a password, enter three characters. The
Horizontal Scroll
bar selects between
the various fields; the
Vertical Scroll
bar changes the value in the selected field. For security
purposes, only the value of the currently selected field is displayed, and the other two fields
5. Press
F1
to apply the new password.
NEW PASSWORD E 10:31
Password 0** ^
< > v
F1=Apply F4=Cancel
Figure 25: Entering a New Password
When password protection is enabled, access to any menu that will result in a change to the stored
system constants will require log in with the password. This restriction includes all menus under the
Mode
,
I/O
, and
Setup
options. Options under
Maintenance
and
Data History
, as well as the
Start
and
Stop
analysis functions, do not require a login.