CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018
98 of 112
DLM 97200-03 EN Rev. A
2. Press
F4
to select the
Maintenance
screen.
3. Press
F1
to select the
Diagnostics
screen. When the
Diagnostics
menu displays, the Sensor
will prompt you to stop analysis.
4. Use the
to scroll down to the
Save Service Data
option, and then press the
button.
5. The data files are saved to the USB flash memory drive. If Technical Support or your local service
provider has asked to review the file, you can take the flash memory drive to your personal
computer and attach the service data file to an email.
The following files are downloaded:
Table 6:
Data Files Saved with “Save Service Data” Option
Example File Name
File Type
SERV_9_2008_03_10_001.csv
1
service data
CON_9_2008_03_10_001.csv
system constants
DATA_9_2008_03_10_001.csv
data history
CAL_9_2008_03_10_001.csv
calibration history
ERR_9_2008_3_10_001.csv
2
error history
1
Depending on the date of download and whether this was the first download of the day, the service data file would be named
as follows: SERV_9_2008_03_10_001.csv.
2
ERR_9_2008_3_10_001.csv, the file name signifies the following: name (error log); serial number; date; and file number for
that day.
Problems Connecting to the Sensor Via Ethernet
If your connection to the Sensor is disrupted or difficult to establish, try the following:
•
Cycle power to the Sensor
•
Confirm that Modbus is enabled (via the I/O menu)
No Flow Conditions
If it appears that sample water is not flowing through the Sensor, verify the no-flow condition by
performing a