4. Make sure the analog output is enabled by highlighting
Analog Output
to toggle the analog
output state between
On
and
Off
.
5. Select
Set Parameter
(TOC, Conductivity/Resistivity, Temperature).
•
Press
F1
to select TOC, press
F2
to select conductivity or resistivity, or press
F3
to
select temperature as the analog output parameter. (You can specify whether data
should be reported as conductivity or resistivity via the
Setup
menu,
Units of
Measure
. Conductivity will be either temperature corrected or raw conductivity as
specified in the
Setup
menu.)
•
The
Analog Min
screen displays, indicating the most recent minimum value for the
selected parameter. To change the value, use the
Horizontal Scroll
bar to select the
field, and use the
Vertical Scroll
bar to change the value in the selected field. Press
F1
to apply the new value.
•
The
Analog Max
screen displays, indicating the most recent maximum value for the
selected parameter. To change the value, use the
Horizontal Scroll
bar to select the
field, and use the
Vertical Scroll
bar to change the value in the selected field. Press
F1
to apply the new value.
6. Analog MIN/MAX is also available without reselecting
Set Parameter
.
7. The Sensor allows you to set the output level when the Sensor switches into Standby mode,
when an error is detected, or when a warning is detected. Select
Standby
to set the Standby
output. Select
Error
to set the error output. Select
Warning
to set the Warning output.
•
Use the Vertical Scroll bar to highlight a value, and then press the
Enter
key to select
that value.
•
The available values are
0.0 mA, 2.5 mA, 4.0 mA, 20 mA
, and
Hold
. The first four
options set the standby or error value to the specified value. The
Hold
option sets the
standby or error value to retain the last value output by the Sensor.
•
Make sure you have set different values for
Standby
,
Error
, and
Warning
.
Binary Input
The CheckPoint can be configured with one remote start binary input, as follows:
The
Binary Input
screen shows the current state of the binary input.
1. Scroll to
Binary Input
and press
.
2. Press
F1
to toggle the binary input state between
Enabled
and
Disabled
. Set the value to
Enabled
to enable remote start to begin the analysis.