CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018
74 of 112
DLM 97200-03 EN Rev. A
When the battery has been drained of power, the battery icon changes to the dead battery warning
icon in the upper right corner (see
The Dead Battery Warning Icon). When analysis is not
running, this icon flashes once per second. When analysis is running, this icon alternates with the
analysis icon once per second.
MAIN
10:31
F1=Mode
F2=I/O
F3=Setup
F4=Maintenance
Figure 30: The Dead Battery Warning Icon
N
OTE
: Only use the CheckPoint with a battery previously provided by SUEZ.
LED Indicator
The CheckPoint LED indicator light illuminates when the Sensor is drawing AC power (when it is
plugged in to a wall outlet or when it is wired to power conduit). The LED will not illuminate when the
Sensor is running on battery power.
Connecting the Vial Sampling Kit
To use CheckPoint to measure from 30-mL vials, you must connect the vial sampling kit as follows:
1. If the Sensor is taking measurements, press the
button and
F1
.
2. Disable the flow of sample water to the Sensor.
3.
Disconnect the sample inlet fitting from the Sensor’s
Inlet
port.
4.
Disconnect the tubing from the Sensor’s
Bypass
port.
5. (Optional) Disconnect th
e tubing from the Sensor’s
Waste
port if you will be connecting waste
tubing directly to the waste container on the vial sampling kit.
6. Loosen the two screws securing the fluidics cover on the right end of the Sensor and connect the
vial sampling kit to the Sensor. Slide the notches in the vial sampling kit bracket over the screws