CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018
97 of 112
DLM 97200-03 EN Rev. A
List of Errors
#
Name
Description
102
Invalid TOC value
Invalid TOC error. TOC reading is negative.
Check UV lamp, recalibrate.
103
System constants write
Problem encountered writing system constants to
flash. If the problem persists you may need to
change the system board. Contact Technical
Support or your local service provider.
104
System constants queue overflow
Problem encountered writing system constants to
flash. If the problem persists you may need to
change the system board. Contact Technical
Support or your local service provider.
105
Battery drawing power on AC
Power supply problem. Battery is showing current
draw while AC is connected.
106
Excessive noise on cell 1
Data rejection on cell 1 exceeds 90%, indicates
problem with cell or conductivity board.
107
Excessive noise on cell 2
Data rejection on cell 2 exceeds 90%, indicates
problem with cell or conductivity board.
108
Analog board communications
Unable to communicate with EEPROM on
conductivity board at power on. Check cable
connections between conductivity board and
system board. If problem persists, replace the
conductivity board.
109 System constant read
Bad record encountered reading the system
constants file. Contact Technical Support or your
local service provider. The most likely cause is
unexpected power outage during write (Shutting
down without using the
Off
button). System
constant will be reset to default value. If this error
occurs frequently, this may indicate problems with
flash on the system board. You will need to
restore constants to correct the problem.
110
Battery over temperature
Battery is reporting over temperature condition.
Check that the fan is running the fan outlet and air
intakes are not blocked.
Step 2: Basic Troubleshooting Steps
Downloading Service Data
Technical Support or your local service provider
may ask to review the Sensor’s service data while
assisting with troubleshooting. To download the service data, make sure a USB flash memory drive is
attached to the Sensor’s USB port (located behind I/O cover) and follow these steps:
1. Press the
button.