14. Legacy
14.1 Suspension Installation
1. Assemble and hang suspension Assembly.
a. If suspension is lights suspension only,
attach down tube to suspension, slide
reducing ring over down tube (set
screws down), then attach flange to
down tube, installing all six screws and
bolts.
2. Place light suspension system securely on
material-lift (e.g., Genie Lift).
3. Use cardboard to protect arm from
scratching on lift.
4. Position material-lift directly under the ceiling mount.
5. Raise material lift up slowly, guiding ceiling mount onto the all-thread rods.
6. Pull all cables and electrical connections through mounting (interface) plate.
7. Using alternating bolts, only secure three hex nuts.
8. Using a Torpedo Level, level the flange to zero across three horizontal planes.
9. Install remaining three flat washers, lock washers, and hex nuts (24mm).
10. Remove original three hex nuts only and replace with three flat washers, lock washers, and hex
nuts (24mm).
11. Tighten all six nuts.
12. Verify that the flange is level across three horizontal planes.
13. Lower the material lift and check light for steadiness.
14.2 Installing Spring Arms (if necessary for lights only suspension)
Top Cover
1. Remove top cover at the end of
the Suspension Arm.