![Stryker MMP200 Installation And Service Manual Download Page 77](http://html.mh-extra.com/html/stryker/mmp200/mmp200_installation-and-service-manual_1385968077.webp)
77
S
Warning
Improper installation may result in the end cap falling into the sterile field.
Back View
Front View
Countersunk Screw (2 for each end cap)
Threaded Hole
Extension Arm End Cap
Slotted Clips (oriented to face down)
Extension Arm Pin (2 for each end cap)
MMP200/OSC600/OSC400 End Cap Installation
12.4.4 Mounting Motor Ring Covers
1. Place the semicircular covers around the
motor arm bearing and fix into place by
screwing in two plastic screws M3 X 8 mm
and two MLF screws M3 X 6 mm
2. Check that the semicircular covers are
securely in place.
12.4.5 Replacing Rear End Spring Arm Covers
1. Use screw driver to separate covers.
2. Place new cover on spring arm.