S
Support Straps
Support strap locations
(each yoke will only contain two support straps
even though four are shown in this figure)
4. Remove the eight M5 screws located under
the Width Adjust Clamp using a 4mm Al-
len wrench, and pull down the clamp.
5. Loosen the support strap inside the Width
Adjust Clamp that is securing the support
tube in place from either the left or the
right side. Do not remove the strap com-
pletely.
Note
For ease of installation, adjust one side at a time.
6. Yoke width is adjusted by inserting the tube alignment pins into the numbered holes in the
bottom piece of the Width Adjust Clamp. Use Table 10.1 to determine the correct holes for the
monitor, and seat the alignment pins in their respective bracket holes.
Note
Measurements presented in the table are estimates only. Adjustments may need
to be made before completing installation to ensure correct balance and fit.
7. Secure the horizontal tube to the lower Width Adjust Clamp by tightening the support straps
with self tapping screws until the strap is torqued. Avoid tearing the strap.
WARNING Failure to tighten the support straps could result in the yoke falling.
Note
Use the inside strap locations when possible.
8. Repeat steps 5-7 for the opposite side.
9. Reassemble the Width Adjust Clamp and secure using the M5 socket head screws. Tighten all
screws to 35 lb. in. (3.95 Nm) torque.
WARNING Ensure that all screws are present. Missing screws could result in the yoke
falling.
Note
Use a torque wrench to tighten screws. If a torque wrench is not available, do not
attempt to tighten screws yourself, but contact your Stryker representative.