Cleaning, Maintenance, and Disposal
Warning
•
Do.not.remove.the.cover.of.the.console.as.this.could.cause.electric.shock.and.product.damage.
•
To avoid electric shock and potentially fatal injury, unplug the pump from the electrical outlet
before cleaning�
Caution
To prevent product damage:
•
when cleaning the pump, do not spray cleaning liquid directly onto the pump; spray on the cloth
before wiping the pump,
•
do not immerse the pump in any liquid,
•
do not use corrosive cleaning solutions to clean the pump,
•
do not sterilize the pump�
Clean the Components
Pump
Should the pump need cleaning:
1� Spray cleaning liquid onto a dry, sterile cloth� Avoid excess liquid or drips�
2� Wipe the pump�
3� Take extra care when cleaning the front LCD screen� Excess liquid or drips that enter the bottom of the screen
may result in product damage�
Footswitch
Warning
Clean the footswitch prior to first use and after every subsequent use�
Consult the footswitch manual (P17862) for cleaning instructions�
Hand control
Warning
Clean and sterilize the hand control prior to first use and after every subsequent use to minimize risk
of infection�
Consult the hand control manual for cleaning and reprocessing instructions�
Replace the Fuses
1� Disconnect the pump from the electrical output and remove the power cord from the rear of the pump�
2� Remove the appropriate fuse holder; with a straight blade screwdriver, unlatch the fuse drawer, or turn the
global fuse holders in a counter clockwise direction until the spring pushes out�
3� Remove the fuse(s)�
4� Replace the fuse(s) with a fuse of the same value and rating as indicated on the rear of the pump�