EN-27
DRAFT
Navigating to the Help Menu
1� Press
to open the Help Menu on any screen�
System Feedback
Audible Feedback
Audible Feedback
Event
Implication
Three high-tone beeps
Fault error
Refer to “Error Messages”
Three high-tone beeps,
pause, two beeps
Lockdown error
Refer to “Error Messages”
One medium-tone
beep
Accessory connection
Refer to “Connect the Components” and “Insert
the Inflow/Outflow Cassette”
Two medium-tone
beeps
Accessory removal
Refer to “Connect the Components” and “Insert
the Inflow/Outflow Cassette”
One low-tone beep
Touchscreen/Hand control/
Footswitch operation
N/A
Three (medium, low,
high-tone) beeps
Hot swap
Refer to “Swap between Hardware Combination
Selections”
Error Messages
The words caution, fault, and lockdown carry special meanings and should be carefully reviewed:
•
Caution:
Visual popup error is displayed for five seconds; user may clear the message when the problem
is resolved�
•
Fault:
The pump will stop; user may clear the message when the problem is resolved�
•
Lockdown:
The pump will stop; reboot system�
Error Message
Cause
Possible Solution
CAUTION
Cassette about to expire (two hours left
before expiration)�
• Replace the cassette�
CAUTION
Preventative Maintenance required; the
unit has reached its recommended service
interval�
• Contact your Stryker
representative�
CAUTION
Inflow/Outflow RFID read failure: The
Inflow or Outflow RFID tag cannot be
detected while the pump is stopped�
• Ensure the Inflow or Outflow
Cassette is fully inserted�
• If the problem persists, replace
the cassette�
• If the problem persists, contact
your Stryker representative�
FAULT
Priming Error: No fluid in the tubing after
ten seconds�
• Ensure the pinch clamps on the
saline bags are open�
• Ensure the Inflow Cassette is
fully inserted�