There are abnormal
pressure or flow rate
fluctuations�
There is a pressure sensor error�
• Reboot the system�
• If the problem persists, contact your
Stryker representative, or return the
console for repair�
A hardware fault is detected�
The pump stops
pumping fluid, and
the pressure indicator
continues to blink�
The pressure is too high� The actual
pressure exceeds 200 mmHg for
15 seconds or greater than 250 mmHg
for five seconds�
• Open the valve at the outflow tube, the
drainage tube, or the stopcock on the
instrument to reduce the pressure�
The pinch valves are
not engaging�
There is a pinch valve or hardware
error�
• Examine the pinch valves to ensure
they are functioning properly�
• Reboot the system�
Bubbles appear in the
joint�
There is no more irrigation fluid�
• Replace or add additional saline bags�
• Ensure the pinch clamp is open on the
irrigation tube, or the stopcock is open�
The tubing connection is loose�
• Ensure the tubing is securly connected�
• If the problem persists, replace the
tubing�
A priming error occurred�
• Ensure the pinch clamps on the saline
bags are open�
• Ensure the Inflow or Outflow Cassette is
fully inserted�
• Remove the hardware from the tubing
and repeat the priming step�
• If the problem persists, replace the
cassette�
The suction level is set too high�
• Decrease the suction level�
The pump cannot
achieve the set
pressure�
There is a hardware setup error, the
flow limit is set too low, or the suction
level is set too high�
• Verify the luer-lock is tightly closed,
the correct hardware is selected and
properly connected, and the dual
stopcock cannula is properly set up�
• Increase the flow limit or the set
pressure setting�
• Decrease the suction level�
• Press Run/Stop to restart the pump�
There is insufficient
pressure in the surgical
site�
An irrigation problem exists�
• Check the stopcock on the arthroscope
and the clamps under the saline bags�
The hardware set up is incorrect�
• Ensure the hardware is properly
selected and set up�
There is no suction
while the pump is
running�
The approved resection consoles are
improperly connected�
• Ensure the consoles and the tubes
are properly connected according to
the instructions in Setup and Device
Configuration section of this manual�
The tubes on the Outflow Cassette are
improperly connected�
Unable to upload files
to/from USB drive�
A hardware fault occurred�
• Ensure the USB drive is functioning
properly and the correct files are on
the drive� If the drive is not functioning
properly, replace the USB drive�
• Reboot the system�