EN-35
DRAFT
• Empty the measuring cup�
• Unclamp the tubing and place it in the cup� Start the stopwatch once the tubing is placed back in
the cup�
• Once the measuring cup is filled with 1 liter of fluid, press
to stop the pump�
• The test has been successfully completed if the time it takes to fill the measuring cup with 1 liter of
fluid is within the range specified in the table below�
Flow (%)
Time (seconds)
50
70
Perform Preventive Maintenance and Calibration
Caution
Stryker does not accept any liability for direct or consequential damages if:
•
the pump or the accessories are improperly prepared and maintained,
•
non-authorized persons perform repairs, adjustments, or alterations to the pump or accessories,
•
non-authorized persons open the pump,
•
the prescribed inspection and maintenance schedules are not followed�
When the pump is 95% through the current maintenance period, a “Service Pump Soon” notification
will appear on the screen� An authorized Stryker service technician must inspect and service the device
according to the maintenance and calibration schedule below to maintain product functionality�
Component
Maintenance/Calibration Period
Motors
2
years
Pinch valves
2
years
Pressure transducer
2
years
Expected Life
Equipment
Expected Life
Console
Five years
Footswitch
Three years
Hand Control
One year
Inflow, outflow, patient-use tubing
Single-use
Day-use cassette
Ten cases, eight hours of active use, or 24 hours after point of first
use
Disposal
This product contains electrical waste or electronic equipment� It must not be disposed of as unsorted
municipal waste and must be collected separately in accordance with applicable national or institutional
related policies relating to obsolete electronic equipment�
Dispose of any system accessories and irrigation fluid according to normal institutional practice relating to