EN-22
DRAFT
Shaver Console Settings
1� Press
to specify the shaver console:
• Press
to select the CrossFire
console;
• Press
to select an approved
resection console;
• Press
if no shaver console is in
use�
In this scenario, the CrossFire console is
selected�
2� Press and to increase or decrease
the Suction (adjust in increments of 10% from
0–100%)�
In this scenario, the Suction is set at 50%�
RF Console Settings
1� Press
to specify the RF console:
• Press
to select the CrossFire
console;
• Press
to select an approved
resection console
(Press “Suction” if the RF probe is a suction
probe�);
• Press
if no RF console is in use�
In this scenario, an approved resection
console with a suction probe is selected�
2� Press and to increase or decrease
the Suction (adjust in increments of 10% from
0–100%)�
In this scenario, the Suction is set at 50%�
Programming the Footswitch and Formula Shaver Settings
1� Press
to open the Main Menu�
2� Press
to open the Footswitch/Formula Shaver Settings Menu�
Setting
Function
Wash
• Inflow-only mode: Increases set pressure and flow limit by user-specified
percentage over user-specified duration�
• Inflow/Outflow mode: Increases set pressure and suction by user-specified
percentage over user-specified duration�
Clear
• Inflow-only mode: Increases flow limit by user-specified percentage over user-
specified duration�
• Inflow/Outflow mode: Increases the suction by user-specified percentage over
user-specified duration�
Drain
• Operates the outflow pump to remove fluid from the surgical site for 30
seconds or until the user stops the pump�
Summary of Contents for CrossFlow
Page 1: ...DRAFT 0450000000 CrossFlow Integrated Arthroscopy Pump DRAFT ...
Page 2: ...DRAFT ...
Page 46: ...EN 44 DRAFT ...
Page 47: ...DRAFT ...