EN-31
DRAFT
Sporadic electrical
interference is affecting
the pump�
Electrical, RF, and/or mobile
communications equipment is
affecting the normal function of the
pump�
• Power down all electrical equipment
not in use�
• Increase the distance of other electrical
equipment�
• Connect the pump and other
equipment into different outlets�
The touch screen
is unresponsive or
inaccurate�
Touch screen is not properly calibrated
for the user�
• Reboot the system�
• If the problem persists, contact your
Stryker representative or return the
console for repair�
A hardware or software error occurred�
A user preference
file was accidentally
deleted from the pump�
Unintentional deletion of a user
preference file�
• Transfer the user preference file from
a USB drive to the pump according to
the instructions in the “Transfer Files
to/from a USB drive” section of this
manual�
The pump is stopped,
but the Outflow
Cassette is stuck and
cannot be ejected�
The pinch valves do not retract�
• Do not attempt to remove the cassette
as it may damage the pump or the
cassette�
• Reboot the system�
• If the problem persists, contact your
Stryker representative�
“Service Pump Soon”
The pump is 95% through the current
maintenance period�
• Contact your Stryker service
representative�