45
Electromagnetic Compatibility
Like other electrical medical equipment, the Precision AC Camera requires
special precautions to ensure electromagnetic compatibility with other
electrical medical devices. To ensure electromagnetic compatibility (EMC),
the camera system must be installed and operated according to the EMC
information provided in this manual.
Note:
The Precision AC Camera has been designed and tested to comply with
IEC 60601-1-2 requirements for EMC with other devices.
• Do not use cables or accessories other than those
provided with the Precision AC Camera, as this may result
in increased electromagnetic emissions or decreased
immunity to such emissions.
• If the Precision AC Camera is used adjacent to or stacked
with other equipment, observe and verify normal operation
of the camera system in the configuration in which it will
be used prior to using it in a surgical procedure. Consult
the tables below for guidance in placing the Precision AC
Camera.
• Equipment which employs RF communications may affect
the normal function of the Precision AC Camera.
Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
The Precision AC Camera is intended for use in the electromagnetic environment specified below.
The customer or user of the camera system should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment - guidance
RF emissions CISPR 11
Group 1
The Precision AC Camera uses RF energy
only for its internal function; therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The Precision AC Camera is suitable for use
in all establishments other than domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes,
provided the following warning is heeded:
Warning
: This system is intended for use by
health care professionals only. This system
may cause radio interference or may disrupt
the operation of nearby equipment. It may be
necessary to take mitigation measures, such as
reorienting or relocating the system or shielding
the location.
Harmonic emissions
IEC61000-3-2
Class A
Voltage Fluctuations/ flicker
emissions IEC61000-3-3
Complies