Performing the White-Balance Test
Before each surgical procedure, perform the white-balance test to adjust the
camera’s perception of white so it can display other colors correctly.
1. Ensure that a scope and light source are attached to the camera system,
and that the console, light source and monitor are powered on.
2. Point the scope tip at several stacked white gauze pads, a white
laparoscopic sponge, or any clean white surface.
3. Look at the monitor and make sure there is no visible glare off of the
white surface of the image.
4. Press and hold the camera head W button (or “WB” on the touchscreen)
until the monitor displays the message “WHITE BALANCE IN PROGRESS.”
5. Continue pointing the scope at the white surface until the video monitor
displays the message “WHITE BALANCE COMPLETE.” The image may
change color. If you cannot achieve an acceptable white balance, refer to
the “Troubleshooting” section.
Advanced Features
The Precision AC Camera has additional features that are not detailed in this
guide:
• Video image settings
• Button programming
• Touchscreen language settings
• Light source “Run/Standby” controls
• Other system settings
These advanced features require in-depth knowledge of the device and
should be performed only by trained personnel. For access to advanced
features, contact a Stryker representative.