38
Sterrad
1. Clean and prepare the camera head and cable as recommended
in this user guide. Ensure the soaking cap is installed. Refer to the
“Installing the Soaking Cap” section for more detail about installing
the cap.
2. If using a sterilization tray (optional), follow any additional
instructions provided with the tray. Use only trays that are compatible
with Sterrad.
Not all sterilization trays are compatible with Sterrad
systems. Using an incompatible tray may result in
incomplete device sterilization. Consult the instructions
that came with your sterilization tray to determine which
sterilization method is compatible with your tray and
devices. If a compatible tray is not available, the devices
can be double wrapped prior to using the Sterrad system.
3. Double wrap the camera head and cable (or tray) prior to sterilization
4. Sterilize the camera head and cable following the instructions of the
manufacturer, using the Sterrad 100S, 200, NX, or 100NX Sterilization
System. Select the standard cycle.