32
3. Rinse
• Remove the device from the prepared detergent. Rinse the device
with reverse osmosis/deionized water at ambient temperature for
90 seconds or until all visible detergent residue has been removed.
• Flush any mated surfaces and crevices a minimum of 5 times. After
all visible detergent residue has been removed, continue to rinse
for 30 seconds.
• Drain excess water from the device and dry it using a clean cloth or
filtered pressurized air.
• Visually inspect the device for cleanliness, paying close attention to
hard-to-reach areas. If visible soil remains, repeat steps 1 through 3.
4. Soak
• Prepare a solution of non-enzymatic detergent and tap water
at 35 °C. Refer to the manufacturer’s instructions for detergent
concentration and any other recommendations.
• Fully immerse the device and use an appropriately-sized syringe to
inject any mated surfaces and crevices with at least 50 mL of the
prepared detergent.
• Soak the device for a minimum of 15 minutes.
5. Brush
• Thoroughly brush the exterior of the device with a soft-bristled
brush for 90 seconds, focusing on any crevices and mated surfaces.
• Inject any mated surfaces and crevices a minimum of 5 times with
50 mL of the prepared detergent.
6. Rinse
• Remove the device from the prepared detergent. Rinse the device
with reverse osmosis/deionized water at ambient temperature for
90 seconds or until all visible detergent residue has been removed.
• Flush any mated surfaces and crevices a minimum of 5 times. After
all visible detergent residue has been removed, continue to rinse
for 30 seconds.
• Drain excess water from the device and dry it using a clean cloth or
filtered pressurized air.
• Visually inspect the device for cleanliness, paying close attention to
hard-to-reach areas. If visible soil remains, repeat steps 1 through 6.