14
Setting Up the Coupler
Steps 1–3 below provide instructions for connecting the camera head to the
coupler.
When using a direct-coupled C-Mount endoscope (a scope that requires no
coupler), thread the endoscope directly into the camera head until it forms a
tight seal, and skip to step 3.
When attaching or removing the coupler, grip only the rear
adapter, as twisting other parts of the coupler with force may
result in mechanical damage.
Do not overtighten the coupler (or a direct-coupled C-mount
scope), as this may damage the front window of the camera.
1. Attach the coupler to the camera head.
• Gripping the rear adapter, screw the coupler clockwise onto the
camera head until it forms a tight seal.
Summary of Contents for 0700010000