33
Automated Cleaning
Caution
If the camera head is cleaned with the automated cleaning
method, steam sterilization is recommended.
1. Soak
• Ensure the soaking cap is installed. Refer to the “Installing the
Soaking Cap” section for more detail about installing the cap.
• Prepare a solution of enzymatic detergent and tap water at
35 °C. Refer to the manufacturer’s instructions for detergent
concentration and any other recommendations.
• Fully immerse the device and use an appropriate-sized syringe to
inject any mated surfaces and crevices with at least 50 mL of the
prepared detergent.
• Soak the device for a minimum of 15 minutes.
2. Brush
• Thoroughly brush the exterior of the device with a soft-bristled
brush for 90 seconds, focusing on any crevices and mated surfaces.
3. Rinse
• Remove the device from the prepared detergent. Rinse the device
with reverse osmosis/deionized water at ambient temperature for
90 seconds or until all visible detergent residue has been removed.
• After all visible detergent residue has been removed, continue to
rinse for 30 seconds.