service manual Safety Information
Stille imagiQ2
TM
service manual Safety Information
n
Ensure that hands, feet and equipment are clear of the frame assemblies when changing positions of the table.
n
Use precaution as the imagiQ2™ table can be activated by another operator.
n
Particular precaution must be taken during use of the foot control to avoid unwanted movements.
n
Not intended to be used in an oxygen rich environment.
n
Do not subject the operating table to electromagnetic fields that exceed the applicable standards for radio
frequency equipment, such as diathermy apparatus, defibrillators and defibrillator monitors.
n
All service shall be performed without any patient placed on the table. In case of emergency service and repair
shall be performed without any connections between the technician and the patient or the operator.
n
For moving safely over a threshold 15 mm or higher, it´s required to be two persons.
n
Before moving over a threshold always makes sure that the table top is centre.
n
The battery shall only be changed by service personnel.
n
Replenishment of oil into the hydraulic system shall only be done by service personnel.
n
Addition to the Service Manual: Never overfill the hydraulic system when refilling oil.
n
Do not place or position imagiQ2 so that it is difficult to disconnect the table from the supply mains socket.
n
Do not touch the patient and accessible contacts of connectors simultaneously.
n
The table shall be centered and turned OFF on the hand control before and during transport.
n
Extra precaution must be taken during service when a guard is open and someone is activating a movement.
Risk of injury.
n
A clicking sound shall be heard when attaching the head rest to be sure that its attached safely.
n
Preventive maintenance of the table shall always be performed according to the Service Manual.
n
Always check that the accessories are securely attached to the table before use.
n
Always check accessories before use. Damaged or worn accessories shall never be used as they can cause
injury.
n
Do not operate the device if the protective covers are removed, risk for personal injury.
n
Skin contact with hydraulic fluid may cause skin irritation, clean and wash with soap and water.
n
If another mattress is used than the original mattress, a grounding connection must be established.
n
When the AC cable is connected to the device, the batteries are charging if the main switch is ON, if the bat-
teries are completely discharged it may take some time before the device can be restarted.
n
Isolation from the supply mains is made by disconnecting the AC power cable from the equipment.
n
Enclosure protection of the foot control is IP68.
No specific warnings specified in the service manual according to use of other accessories, transducers and
cable other than those specified by Stille.
n
Precautions has to be taken to equipment and surrounding equipment in relation to specific EMC requirements
see tables for EMC specifications.
n
Testing against IEC/EN 62353 shall be performed by Qualified personnel according to the requirements in the
standard.
n
Use of accessories, cables and spare parts other than those sold by Stille as replacement for internal compo-
nents, may result in increased emissions or decreased immunity of the operating table imagiQ2.
n
Warning: imagiQ2™ should not be used adjacent to or stacked with other equipment. If adjacent or stacked use
is necessary, imagiQ2™ should be observed to verify normal operation in the configuration in which it will be
used.
n
WARNING: if the table is beeping continously every 30 second, table is in failstate, all safetyfunctions are
turned OFF, do not transport patients on the table park the table during procedures.
n
WARNING: Risk for pinch/squees damage when installing/ removing the detachable table top.
n
WARNING: To avoid the risk of electric shock, this equipment must only be connected to a supply main with
protective earth.
n
WARNING: Using other cables and accessories than the specified, can possibly increase the emission and
decrease the immunity in regards of EMC requirement.
n
WARNING: To avoid risk of electric shock, this equipment must only be connected to a supply mains
with protective earth.
n
WARNING:
No modification of this equipment is allowed.
n
Do not position the imagiQ2™ operation table so that it is difficult to operate the disconnection device.
n
Means for isolating imagiQ2™ operation table from mains supply, trough disconnecting the power cord from
the ME equipment.
n
It is recommended that at least 2 persons are transporting the imagiQ2™ operation table.
n
If in doubt, consult the supplier or person responsible for operating room equipment.