43
Stille imagiQ2
TM
service manual Technical
Technic
al
TROUBLE SHOOTING GUIDE
Electrical problems, non functional table
Symptom
Probable cause
Procedure
The table does not function
while under battery power
but functions normally when
mains power is connected.
a. The battery package is
completely discharged.
b. The battery package is not
correctly connected in the
table.
c. The battery fuse has blown
d. The battery package is
unserviceable/failure in a cell.
a. Charge the batteries
(approx. 10 hours).
b. Check the connections of
the battery package.
c. Check first all the heavy
red/black cables in the table
for damaged insulation/a
short circuit. Check the bat-
tery fuse.
(replace any ”blown fuses”
with new fuses of the same
rating.)
d. Replace the battery pack-
age.