26
Stille imagiQ2
TM
service manual Technical
Technic
al
To replace AC-inlet
1)
If possible place the table in position for service according to instruction, mini-
mum try to elevate the table top to the highest position, lock the wheels.
2)
Remove the emergency cover and then the protective plates under the covers.
3)
Shut down the table on the main switch.
4)
Perform an electric LoTo according to the instruction to remove any stored energy
in the system.
5)
Disconect the AC-inlets three cables. Make a note for the positioning of the cables
for re-attachment.
6)
Unmount the AC-inlet by losening the two screws on the front.
7)
Before mounting the new AC-inlet check that the fuses are of the right ampere.
See lable for fuse rating.
8)
Mount the new AC-inlet.
9)
Thighten the two screws in the front.
10)
Reattached the cables at the AC-inlet, make sure you get the cables positioning
right.
11)
Reinstall the protective covers.
12)
Complete the LoTo procedure