service manual Technical
Technic
al
TROUBLE SHOOTING GUIDE
Hydraulic problems floor locks parking cylinders
Symptom
Probable cause
Procedure
The table can be parked but
the pump motor continues
for approx. 15 seconds. and
an error indication (alterna-
tely blue/green) flashes on
the hand control after the
motor has stopped. Panning
does not function.
Connector (PX) on the main
board is not correctly con-
nected.
a. Check the connector.
The blue/green lamps with
the symbols for locked/open
lock begin to flash alterna-
tely after the table has been
parked a time.
a. Incorrectly adjusted posi-
tion switches
b. Leak in the hydraulic
parking circuit.
c. position switches unservi-
ceable.
a. Adjust the position swit-
ches.
b. Trace any leak in the
hydraulic parking circuit.
c. Replace the position swit-
ches.
The pump motor runs for
approximately 15 seconds
after an attempt has been
made to park the table or
adjust the lateral tilt/set the
zero position, and a fault indi-
cation for parking (alterna-
tely blue/green) is presented
after the motor stops. Pan-
ning is possible.
a. Cable not correctly con-
nected to the position swit-
ches.
b. Incorrectly adjusted posi-
tion switches.
c. Leak in the hydraulic par-
king circuit.
d. Position switches not
giving signal.
a. Check the cable connec-
tions.
b. Adjust the position swit-
ches.
c. Trace any leak in the
hydraulic parking circuit.
d. Replace the position swit-
ches.