TROUBLE SHOOTING GUIDE
Electrical problems, indications on the hand control unit
Symptom
Probable cause
Procedure
Blue ”Charging in progress”
lamp on the hand control
does not illuminate when the
mains voltage power cable is
connected.
First withdraw the plug from
the wall socket and inspect
the mains voltage power
cable. A damaged cable can
become the cause of injury or
death.
a. The cable is not correctly
connected
b. The main power switch on
the table is switched off.
c. There is no voltage at the
wall socket.
d. The primary fuse (s) on the
table has (have) blown. .
e. The signal cable from the
battery charger is not cor-
rectly connected to the con-
trol unit, (PX).
f. The battery charger is
unserviceable
g. The control unit is unservi-
ceable.
A damaged cable must be
replaced immediately. Use a
Stille original cable only.
a. Check the cable connection
b. Check the main power
switch.
c. Check that there is voltage
at the wall socket.
d. Check the primary fuse(s)
on the table.
N.B. The fuse values for
100/ 120/ 230 V are dif-
ferent.
e. Check the connection of
the signal cable. (PX).
f. Replace the main bord.
g. Replace the control unit.