j) Decontamination
Material necessary for the decontamination procedure:
o
Gloves;
o
Glasses;
o
Mask;
o
Laboratory coat;
o
Hydroalcoholic disinfectant solution for device surfaces commonly used for biological
and medical devices (Ex. Phagospr
ay). Decontaminant solution for device’s surfaces
used for biological and medical devices, specifically targeting nucleases and DNA
contaminations (Ex. RNAse away).
Note: Bleach is not recommended; if its use is unavoidable, ensure abundantly rinsing with
water after the bleach treatment
.
o
Wipes.
For the detailed procedure for decontamination of the Geode, please view the
Decontamination Protocol available here:
https://www.stillatechnologies.com/technical-
resources/naica-system-prism3/
k) Disposal
•
Disposal of naica
®
equipment
The disposal of the Geode and the pressure source at the end of the product's life should
comply with the current legislation in force in the country of use regarding electrical and
electronic waste.
l) Shipment
If shipping the Geode or the compressed air system (ISPTLAPG3) is required, use all the
original packaging and cables provided upon reception of the naica® system.
No chips (Sapphire chip/ Opal chip) must be left inside the Geode.
The original double packaging and protective foam should be used for the Geode.
To prepare the Geode for shipment, proceed carefully as follows:
(1) Remove any chip that may remain in the instrument.
(2) Switch off the power switch located at the back of the instrument and remove the power
supply cable.
(3) Switch off the pressure source and remove the pressure tube from the pressure inlet
located behind the instrument, by pushing on the ring-shaped push-in connector while pulling
on the tube.
(4) Remove any chip that may remain in the instrument.
(5) Repack the instrument into the original packaging in the same way it was delivered; include
the power supply cable in the packaging. For detailed packing instructions, please refer to the
contact the Technical Support.
For detailed packaging instructions, please refer to the packaging procedures available here:
https://www.stillatechnologies.com/technical-resources/naica-system-prism3/