Geode H14000 User Manual
26
Figure 12: Live graph of measured pressure and temperature during a run.
5) When the run is done, a pop-up window is
displayed with the message “PCR completed
successfully, touch the screen to continue”. Touch the screen to close it.
6) Open the lid, take out the chips and pull down the lid.
If support is required for the quality of a run, exporting the live temperature and pressure profiles
at the end of a run would be necessary for optimal investigation.
Pause a program
1) During a run, press the
“pause icon” to interrupt the Crystal Digital PCR
TM
temporarily.
2) Confirm the request in the appearing window and the program will be paused.
3) Press the
“play icon”
to continue the program.
Abort a program
1) During a run, press the
“stop icon”
to abort a run. After confirming the request in the
appearing window, two possibilities are accessible (
Fig. 13
): “STOP Experiment” (longer but
preserves droplet crystal) or “KILL Experiment” (shorter but destroys droplet crystal).
Figure 13: Stopping a run.