Geode H14000 User Manual
17
Figure 4: Magnetic clip-on cushions can be easily clipped next to the existing pressure points.
6) Following, ensure the respective Opal chip magnetic frame is placed on the thermal plate;
chamfer placed on the lower left edge.
Place the respective chips in position using the respective magnetic chip frame on the
thermal plate. The correct orientation of the respective magnetic frames is supported by a
blue (left) and white (white) positioning indicator, both on the magnetic frames as well as on
the Geode instrument thermal plate framing (Fig. 5).
Figure 5: Positioning of the Sapphire chip (left), and Opal chip (right) in the respective magnetic
frame on the thermal plate of the Geode instrument.
7) Close the lid of the Geode by pulling the black handle upward. Ensure that the gasket remains
in place ensuring a hermetic seal while closing the Geode lid.
Figure 6: Opening/closing the lid of the Geode.
Magnetic Opal chip clip-on cushions