b) Safety instructions for the Geode
Before the first use of the naica
®
system, please read the User Manual entirely. Special notice
should be taken of the following.
o
Safety-Disclaimer for the Geode:
Caution of dangerous voltage
Please ensure that the voltage indicated on the back of the device exactly
matches the local electrical supply.
Caution of dangerous explosive material
Explosive or reactive material must not be heated or placed under
pressure in the Geode.
Caution- liquids
Ensure that no liquids can enter the device. Sample must be loaded
outside of the Geode.
Caution- environment
The ventilation of the device must not be covered.
Caution- use
If
the device is used in a manner not described in this manual, safety may
be compromised.
Caution- hot surface
The thermal plate and chips quickly reach temperatures above 50°C thus,
there are risks of burning or scalding if the equipment is not operated
properly. Keep the lid closed until the temperature reaches 30°C or less.
Only use materials (chips and caps) provided by Stilla Technologies,
which are heat resistant at temperatures up to 95°C.
Caution- the risk of pinching
The lid of the Geode should be closed with caution before starting a run.
Do not try to open the Geode lid during a run. Do not manipulate the
Geode lid handle during a run. Always check that the internal pressure of
the chamber is below 20 mbars before opening the lid.