985-006001 REV. G Effective Date: 03 June 2008
Stereotaxis, Inc.
Page 14 of 43
8
Verify the system will move to Retracted position from the Navigate
position.
•
While the Positioner is in the Navigate position press and
hold the Retracted position button:
and
on the tableside controller until the Positioner reaches the
Retracted position.
•
Verify that the Retracted Position indicator:
is displayed on the control room and
procedure room monitors.
•
Verify that while the Positioner is moving the green indicator
for the location the Positioner has been requested to move to
blinks during the move.
P F
P F
P F
9
•
Now simultaneously hold the Image Transfer and the
appropriate Movement button to initiate the interlock override
condition and verify that the system will begin to move with
the Stow position button pressed.
+
P F