985-006001 REV. G Effective Date: 03 June 2008
Stereotaxis, Inc.
Page 12 of 43
5
Verify the system will move to Stowed position from Navigate.
•
While the Positioner is in the Navigate position press and
hold the stowed position button:
on the tableside controller until the Positioner
reaches the stowed position.
•
Verify that the Stowed Position indicator:
is displayed on the control room and
procedure room monitors.
•
Verify that while the Positioner is moving the green indicator
for the location the Positioner has been requested to move to
blinks during the move.
•
Verify that the total time to move from Navigate to Stowed
does not exceed 30 seconds.
•
Verify that the first 20
o
of movement, on Pivot axis, towards
stowed does not exceed 15 seconds.
P F
P F
P F
P F
P F