PKG08806-UMDFG 11
ENGLISH
3.2.1
PRIMARY DISPLAY
The center of the display is used to display the primary result(s). This
result is displayed in a large format with a resolution of 5,000:1. When
tolerance is used, the displayed results will show in RED when the result
is "out-of-tolerance". Results "in-tolerance" are displayed in BLACK.
3.2.2
SECONDARY RESULT
The upper right display area shows secondary results. These results
vary depending on the force gage mode and the primary result being
displayed. For example, if the primary result is a peak measurement, the
secondary result displays the real time load being applied.
3.2.3 DIRECTION
Symbols are used to indicate the direction of load for the current readings.
3.2.4 TOLERANCE
When a tolerance is setup for your result, results that are "within" the
tolerance range are displayed in BLACK. If the results is "outside" the
tolerance range, the result is displayed in RED. The gage may also be
setup to give a sound if the result is "outside" the tolerance range. The
sound is configured in SETUP and may be a "beep", "Chirp" or "Tone".
3.2.5 BARGRAPH
The bargraph shows the measured load as detected by the load cell
sensor. The load is displayed from a center line.
Compressive loads fill to the left while tensile loads fill to the right.
Color is used to depict the load status relative to the full scale of the load cell.
The bargraph will display GREEN when the measured load is from 1 to
75% of the full scale capacity.
The bargraph will display YELLOW when the measured load is from
76% to 90%.
The bargraph will display RED when the measured load exceeds 90%.
If the measured load exceeds the full scale capacity of the load cell, the
primary display will show a RED "- - - -", indicating an OVERLOAD condition.
The bargraph also uses markers to show maximum values achieved during
a test. For example, a marker may depict a peak load measurement.
lbf
5.71
5.66
4
Tensile Direction
Display Formats
Primary Display
Secondary Display
Real Time
Peak Measurement
Peak Tension
Real Time Measurement
Peak Compression
Real Time Measurement
Out-of-Tolerance Result
lbf
46.83
lbf
59.71
0.05
84.93
lbf
lbf
112.40
91.85
lbf
lbf
112.40
91.85
104.22
lbf
lbf
112.40
91.85
Summary of Contents for PKG08806-UMDFG