5-5
Remote Programming
INPUT and FILTER COMMANDS
ISRC (?) {i}
The ISRC command sets or queries the input configuration. The parame-
ter i selects A (i=0), A-B (i=1), I (1 M
Ω
) (i=2) or I (100 M
Ω
) (i=3).
Changing the current gain does not change the instrument sensitivity.
Sensitivities above 10 nA require a current gain of 1 M
Ω
. Sensitivities
between 20 nA and 1 µA automatically select the 1 M
Ω
current gain. At
sensitivities below 20 nA, changing the sensitivity does not change the
current gain.
IGND (?) {i}
The IGND command sets or queries the input shield grounding. The
parameter i selects Float (i=0) or Ground (i=1).
ICPL (?) {i}
The ICPL command sets or queries the input coupling. The parameter i
selects AC (i=0) or DC (i=1).
ILIN (?) {i}
The ILIN command sets or queries the input line notch filter status. The
parameter i selects Out or no filters (i=0), Line notch in (i=1), 2xLine
notch in (i=2) or Both notch filters in (i=3).
Summary of Contents for SR830
Page 5: ...1 4...
Page 11: ...SR830 DSP Lock In Amplifier 1 10...
Page 13: ...2 2 Getting Started...
Page 17: ...2 6 The Basic Lock in...
Page 23: ...2 12 Outputs Offsets and Expands...
Page 25: ...2 14 Storing and Recalling Setups...
Page 31: ...3 4 SR830 Basics...
Page 33: ...3 6 SR830 Basics...
Page 37: ...3 10 SR830 Basics...
Page 53: ...3 26 SR830 Basics...
Page 74: ......
Page 83: ...4 30 Rear Panel...
Page 107: ...5 24 Remote Programming...
Page 113: ...5 30 Remote Programming...
Page 117: ...5 34 Remote Programming...
Page 121: ...6 4 Performance Tests...
Page 123: ...6 6 Performance Tests...
Page 125: ...6 8 Performance Tests...
Page 129: ...6 12 Performance Tests...
Page 131: ...6 14 Performance Tests...
Page 133: ...6 16 Performance Tests...
Page 139: ...6 22 Performance Tests...
Page 145: ...7 2 Circuit Description...