Remote Programming
SETUP COMMANDS
OUTX (?) {i}
The OUTX command sets the output interface to RS232 (i=0) or GPIB
(i=1). The OUTX i command should be sent before any query com-
mands to direct the responses to the interface in use.
OVRM i
In general, every GPIB interface command will put the SR830 into the
REMOTE state with the front panel deactivated. To defeat this feature,
use the OVRM 1 command to overide the GPIB remote. In this mode, the
front panel is not locked out when the unit is in the REMOTE state. The
OVRM 0 command returns the unit to normal remote operation.
KCLK (?) {i}
The KCLK command sets or queries the key click On (i=1) or Off (i=0)
state.
ALRM (?) {i}
The ALRM command sets or queries the alarm On (i=1) or Off (i=0)
state.
SSET i
The SSET i command saves the lock-in setup in setting buffer i (1
≤
i
≤
9).
The setting buffers are retained when the power is turned off.
RSET i
The RSET i command recalls the lock-in setup from setting buffer i
(1
≤
i
≤
9). Interface parameters are not changed when a setting buffer is
recalled with the RSET command. If setting i has not been saved prior to
the RSET i command, then an error will result.
Summary of Contents for SR830
Page 5: ...1 4...
Page 11: ...SR830 DSP Lock In Amplifier 1 10...
Page 13: ...2 2 Getting Started...
Page 17: ...2 6 The Basic Lock in...
Page 23: ...2 12 Outputs Offsets and Expands...
Page 25: ...2 14 Storing and Recalling Setups...
Page 31: ...3 4 SR830 Basics...
Page 33: ...3 6 SR830 Basics...
Page 37: ...3 10 SR830 Basics...
Page 53: ...3 26 SR830 Basics...
Page 74: ......
Page 83: ...4 30 Rear Panel...
Page 107: ...5 24 Remote Programming...
Page 113: ...5 30 Remote Programming...
Page 117: ...5 34 Remote Programming...
Page 121: ...6 4 Performance Tests...
Page 123: ...6 6 Performance Tests...
Page 125: ...6 8 Performance Tests...
Page 129: ...6 12 Performance Tests...
Page 131: ...6 14 Performance Tests...
Page 133: ...6 16 Performance Tests...
Page 139: ...6 22 Performance Tests...
Page 145: ...7 2 Circuit Description...