AUX INPUT and OUTPUT COMMANDS
OAUX? i
The OAUX? command queries the Aux Input values. The parameter i
selects an Aux Input (1, 2, 3 or 4) and is required. The Aux Input voltages
are returned as ASCII strings with units of Volts. The resolution is
1/3 mV. This command is a query only command.
AUXV (?) i {, x}
The AUXV command sets or queries the Aux Output voltage when the
output. The parameter i selects an Aux Output (1, 2, 3 or 4) and is
required. The parameter x is the output voltage (real number of Volts)
and is limited to -10.500
≤
x
≤
10.500. The output voltage will be set to
the nearest mV.