4-17
Front Panel
[Expand]
Pressing this key selects the Y Expand. The expand can be 1 (no
expand), 10 or 100. If the expand is 10 or 100, the Expand indicator
below the display will turn on. The output can never exceed full scale
when expanded. For example, if an output is 10% of full scale, the
largest expand (with no offset) which does not overload is 10. An output
expanded beyond full scale will be overloaded.
Short Time Constant Limitations
A short time constant places a limit on the total amount of DC gain
(reserve plus expand) available. If the time constant is short, the filter
slope low and the dynamic reserve high, then increasing the expand may
change the time constant. See the table of time constants and DC gains
in the Gain and Time Constant section.
Summary of Contents for SR830
Page 5: ...1 4...
Page 11: ...SR830 DSP Lock In Amplifier 1 10...
Page 13: ...2 2 Getting Started...
Page 17: ...2 6 The Basic Lock in...
Page 23: ...2 12 Outputs Offsets and Expands...
Page 25: ...2 14 Storing and Recalling Setups...
Page 31: ...3 4 SR830 Basics...
Page 33: ...3 6 SR830 Basics...
Page 37: ...3 10 SR830 Basics...
Page 53: ...3 26 SR830 Basics...
Page 74: ......
Page 83: ...4 30 Rear Panel...
Page 107: ...5 24 Remote Programming...
Page 113: ...5 30 Remote Programming...
Page 117: ...5 34 Remote Programming...
Page 121: ...6 4 Performance Tests...
Page 123: ...6 6 Performance Tests...
Page 125: ...6 8 Performance Tests...
Page 129: ...6 12 Performance Tests...
Page 131: ...6 14 Performance Tests...
Page 133: ...6 16 Performance Tests...
Page 139: ...6 22 Performance Tests...
Page 145: ...7 2 Circuit Description...