4-19
Front Panel
[
+90°] and [-90°]
The [+90°] and [-90°] keys add or subtract 90.000° from the reference
phase shift. The phase does not need to be displayed to use these keys.
Zero Phase
Pressing the [+90°] and [-90°] keys together will set the reference phase
shift to 0.00°.
[Freq]
Pressing this key displays the reference frequency in the Reference dis-
play.
If the reference mode is external, then the measured reference frequency
is displayed. The knob does nothing in this case. If the harmonic number
is greater than 1 and the external reference goes above 102 kHz/N
where N is the harmonic number, then the harmonic number is reset to 1.
The reference will always track the external reference signal.
If the reference mode is internal, then the internal oscillator frequency is
displayed. The oscillator frequency may adjusted with the knob. The fre-
quency has 4 1/2 digits or 0.1 mHz resolution, whichever is larger. The
frequency can range from 0.001 Hz to 102.00 kHz. The upper limit is
decreased if the harmonic number is greater than 1. In this case, the
upper limit is 102 kHz/N where N is the harmonic number.
[Ampl]
Pressing this key displays the Sine Output Amplitude in the Reference
display. Use the knob to adjust the amplitude from 4 mVrms to 5 Vrms
with 2 mV resolution. The output impedance of the Sine Out is 50
Ω
. If the
signal is terminated in 50
Ω
, the amplitude will be half of the programmed
value.
When the reference mode is internal, this is the excitation source provid-
ed by the SR830. When an external reference is used, this sine output
provides a sine wave phase locked to the external reference.
The rear panel TTL Output provides a TTL square wave at the reference
frequency. This square wave is generated by discriminating the zero
crossings of the sine output. This signal can provide a trigger or sync
signal to the experiment when the internal reference source is used. This
signal is also available when the reference is externally provided. In this
case, the TTL Output is phase locked to the external reference.
[Harm #]
The SR830 can detect signals at harmonics of the reference frequency.
The SR830 multiplies the input signal with digital sine waves at a multiple
of the reference. Only signals at this harmonic will be detected. Signals
at the original reference frequency are not detected and are attenuated
as if they were noise.
Whenever the harmonic detect number is greater than 1, the HARM#
indicator in the Reference display will flash to remind you that the
SR830 is detecting signals at a multiple of the reference frequency.
Always check the harmonic detect number before making any
measurements.
If the harmonic number is set to N, then the internal reference fre-
Summary of Contents for SR830
Page 5: ...1 4...
Page 11: ...SR830 DSP Lock In Amplifier 1 10...
Page 13: ...2 2 Getting Started...
Page 17: ...2 6 The Basic Lock in...
Page 23: ...2 12 Outputs Offsets and Expands...
Page 25: ...2 14 Storing and Recalling Setups...
Page 31: ...3 4 SR830 Basics...
Page 33: ...3 6 SR830 Basics...
Page 37: ...3 10 SR830 Basics...
Page 53: ...3 26 SR830 Basics...
Page 74: ......
Page 83: ...4 30 Rear Panel...
Page 107: ...5 24 Remote Programming...
Page 113: ...5 30 Remote Programming...
Page 117: ...5 34 Remote Programming...
Page 121: ...6 4 Performance Tests...
Page 123: ...6 6 Performance Tests...
Page 125: ...6 8 Performance Tests...
Page 129: ...6 12 Performance Tests...
Page 131: ...6 14 Performance Tests...
Page 133: ...6 16 Performance Tests...
Page 139: ...6 22 Performance Tests...
Page 145: ...7 2 Circuit Description...