6-7
Performance Tests
3. Common Mode Rejection
This test measures the common mode rejection of the lock-in.
Setup
We will use the internal oscillator sine output to provide the signal.
Connect the Sine Out to both the A and B inputs of the lock-in. Use equal length cables from A and B to a
BNC TEE. Connect the cable from the Sine Out to the TEE. Do not use any termination.
Procedure
1) {PRESET} (Turn the lock-in off and on with the [Setup] key pressed)
2) Press the keys in the following sequence:
[Freq]
Use the knob to adjust the frequency to 100.0 Hz.
[Channel 1 Display]
Set the Channel 1 display to R.
3) The value of R should be 1.000 V (within 2%).
4) Press
[Couple]
Select DC coupling.
[Input]
Select A-B.
[Sensitivity Down]
Set the sensitivity to 200 µV.
5) Record the value of R.
6) This completes the CMRR measurement test. The common mode rejection is 20log(1.0/R) where R is
in Volts. Enter the results of this test in the test record at the end of this section.
Summary of Contents for SR830
Page 5: ...1 4...
Page 11: ...SR830 DSP Lock In Amplifier 1 10...
Page 13: ...2 2 Getting Started...
Page 17: ...2 6 The Basic Lock in...
Page 23: ...2 12 Outputs Offsets and Expands...
Page 25: ...2 14 Storing and Recalling Setups...
Page 31: ...3 4 SR830 Basics...
Page 33: ...3 6 SR830 Basics...
Page 37: ...3 10 SR830 Basics...
Page 53: ...3 26 SR830 Basics...
Page 74: ......
Page 83: ...4 30 Rear Panel...
Page 107: ...5 24 Remote Programming...
Page 113: ...5 30 Remote Programming...
Page 117: ...5 34 Remote Programming...
Page 121: ...6 4 Performance Tests...
Page 123: ...6 6 Performance Tests...
Page 125: ...6 8 Performance Tests...
Page 129: ...6 12 Performance Tests...
Page 131: ...6 14 Performance Tests...
Page 133: ...6 16 Performance Tests...
Page 139: ...6 22 Performance Tests...
Page 145: ...7 2 Circuit Description...