4-18 Calibration
DS335 Synthesized Function Generator
The last table of 160 words contains sync symmetry values as a function of frequency when sines or squares
are selected. There are three values which precede the table: these are used to control the sync duty cycle
for triangles, ramps, and noise.
392
TRI_SYM
Triangle sync symmetry.0 to 4095. Nominal 2048.
393
RAMP_SYM
Ramp sync symmetry. 0 to 4095. Nominal 2048.
394
NOISE_SYM
Noise sync duty. 0 to 4095. Nominal 2048
395
SYM_0
0 to 4095. Nominal 2048. For dc to 19530 Hz.
396
SYM_1
0 to 4095. Nominal 2048. For 19,531 to 39,062 Hz.
397
SYM_2
0 to 4095. Nominal 2048. For 39,063 to 58,593 Hz.
:
:
:
554
SYM_159
0 to 4095. Nominal 2048. For 3105468 to 3125000 Hz.
Summary of Contents for DS335
Page 2: ...DS335 Synthesized Function Generator...
Page 6: ...iv SRS Symbols DS335 Synthesized Function Generator...
Page 10: ...viii Specifications DS335 Synthesized Function Generator...
Page 18: ...2 4 Introduction DS335 Synthesized Function Generator...
Page 22: ...2 8 Features DS335 Synthesized Function Generator...
Page 26: ...2 12 Function Setting DS335 Synthesized Function Generator...
Page 30: ...2 16 Sweeps FSK DS335 Synthesized Function Generator...
Page 42: ...3 10 Programming Commands DS335 Synthesized Function Generator...
Page 46: ...3 14 Programming Examples DS335 Synthesized Function Generator...
Page 50: ...4 4 Troubleshooting DS335 Synthesized Function Generator...
Page 74: ...5 8 Circuitry DS335 Synthesized Function Generator...