4-12 Performance Tests
DS335 Synthesized Function Generator
2) Set the time interval counter to measure the positive width of the A input.
Record the reading.
3) Set the time interval counter to measure the negative width of the A
input. This reading should be equal to the reading in step 2 < ±13 ns.
Record the result.
THIS COMPLETES THE PERFORMANCE TESTS