Programming Commands 3-9
DS335 Synthesized Function Generator
6
URQ
Set by any key press
7
PON
Set by power on
This status byte is defined by IEEE-488.2 (1987) and is used primarily to report errors in commands received
over the communications interfaces. The bits in this register stay set once set and are cleared by reading
them or by the *CLS command.
DDS Status Byte:
bit name
usage
0
Trig'd
Set when a sweep is triggered.
1
Trig Error
Set when a trigger rate error occurs.
2
Unused
3
Unused
4
Warmup
Set after the warmup period has expired.
5
Test Error
Set if a self test error occurs.
6
Unused
7
mem err
the stored setting were corrupt on power up.
The Warmup bit will be set and remain set after the warmup period has expired. The rest of the bits in this
register are set when the corresponding event occurs and remain set until cleared by reading this status byte
or by the *CLS command.
Summary of Contents for DS335
Page 2: ...DS335 Synthesized Function Generator...
Page 6: ...iv SRS Symbols DS335 Synthesized Function Generator...
Page 10: ...viii Specifications DS335 Synthesized Function Generator...
Page 18: ...2 4 Introduction DS335 Synthesized Function Generator...
Page 22: ...2 8 Features DS335 Synthesized Function Generator...
Page 26: ...2 12 Function Setting DS335 Synthesized Function Generator...
Page 30: ...2 16 Sweeps FSK DS335 Synthesized Function Generator...
Page 42: ...3 10 Programming Commands DS335 Synthesized Function Generator...
Page 46: ...3 14 Programming Examples DS335 Synthesized Function Generator...
Page 50: ...4 4 Troubleshooting DS335 Synthesized Function Generator...
Page 74: ...5 8 Circuitry DS335 Synthesized Function Generator...