LiQuilaz II E and S Operations Manual
Page E-1
Appendix E
Configuring for Ethernet
Configuration Basics
Configuration is the process of entering Ethernet settings that will allow the LiQuilaz II
Liquid Particle Counter to communicate over a specific network.
Appropriate Ethernet settings are required for both the PMS Ethernet protocol, used
with Facility Net, and the Modbus TCP protocol (see
Protocol”) for use with third party applications.
NOTE:
The following information assumes that the user has a working
knowledge of terminal emulation software such as
HyperTerminal
®
.
When to Configure the Sensor
The sensor should have the appropriate Ethernet settings configured
before
it is wired
to a network. This can be performed either before fluidic installation, if access to the
back of the sensor is problematic, or in-situ, if back panel access is not an issue.
Required Hardware and Software
•
RS-232 cable and USB to RS-232 adaptor (if the PC lacks a serial port)
•
Computer loaded with terminal emulation software
NOTE:
Microsoft Windows® includes a terminal emulator called
HyperTerminal. Other terminal emulator programs, are widely
available. Your IT administrator or microcomputer support
person may be able to provide assistance.