Appendix D Ethernet Communications Overview
LiQuilaz II E and S Operations Manual
Page D-9
Figure D-5 Internet Protocol (TCP/IP) Properties
window
d.
Record the existing settings on the
Internet Protocol (TCP/IP) Properties
window.
These settings will need to be restored after the instrument communication
session is complete.
e.
Change the settings on the
Internet Protocol (TCP/IP) Properties
window to
use the network address and mask determined above. See
-9 for an example. The
Default gateway
and
DNS server
values should be left
blank.
f.
Click the
OK
button to close the
Internet Protocol (TCP/IP) Properties
window.
g.
Click the
OK
button to close the
Local Area Connection Properties
window.
4.
The PC is now configured.
5.
Attach a network cable between the PC and the instrument.
6.
Use the procedure
Troubleshooting Ethernet Connectivity
D-4 to check
the connection is functioning.