Page 3-6
LiQuilaz II E and S Operations Manual
Chapter 3 Maintenance
2.
Carefully remove the top locking pin using a pair of needle nose pliers.
Figure 3-5
Locking pin removal
3.
Use the 3-prong wrench (P/N 1000014376, supplied in the ship kit) to unscrew the
fitting from the top capillary mount.
Figure 3-6
3-prong wrench placement