28
3. Press the Salinity—Product Button
4. Press Continue to acknowledge the
warning
Salinity Calibration
1. Press the Menu Button
2. Press the Calibrate Sensors Button
The Salinity probe has been calibrated at the factory during testing and is not normally
required during commissioning. If the product quality is not reading accurately, follow
calibration steps.
A handheld salinity meter (or other reliable device) is required to perform this calibra-
tion as you need to confirm the salinity of the product water.
5. Allow the salinity to stabilize for 5
minutes.
6. Press the PPM field and enter the
PPM you measured. Press Proceed to
save your entry.