ST2515 Operation Manual
37
SENSE HI: Voltage-sense high terminal SENSE LO: Voltage-sense low terminal
5.1.7
Time Expenditure of Measurement
Instruction
Time
Offset voltage compensation
(OVC) OFF
Offset
voltage
compensation
(OVC) ON
Error test signal output
response
100µs
100µs
Trigger pulse width
100µs
100µs
Measuring delay
Auto, manual setup
Auto, manual setup
Measuring
time
FAST
5ms
10ms+measuring delay
MED
20ms (50Hz)
16.7ms (60Hz)
40ms + measuring delay (50Hz)
33.3ms + measuring delay (60Hz)
SLOW1
100ms
200ms + measuring delay
SLOW2
400ms
800ms + 7 × measuring delay
Computing
time
FAST
1ms
1ms
MED
1ms
1ms
SLOW1, 2
1ms
1ms
Loading time
70ms min, 190ms max
70ms min, 190ms max
5.1.8
Average
Range from 1 to 255, programmable: this value reflects the measurement times from measuring
resistance to measuring display. The setup of average is shown in the average settings of
<Measurement Setup> interface.
5.1.9
Display Digits
SLOW2, SLOW1, MED
FAST
Resistance range
6 digits, maximum display: 110M
5 digits, maximum display: 110M
Temperature range
3 digits, maximum display
number: 999.9°C
3 digits, maximum display